All models of the PSB, PSI, and ELR 10000 series feature an IU table function as part of the function generator. The IU table function (XY table) offers the user the ability to set a DC output current based on the DC voltage measured at the DC terminals. This feature works in either source or sink mode. The IU table function can be used to simulate
Introduction
All models of the PSB, PSI, and ELR 10000 series feature an IU table function as part of the function generator. The IU table function (XY table) offers the user the ability to set a DC output current based on the DC voltage measured at the DC terminals. This feature works in either source or sink mode. In the IU function, an internal measuring circuit measures the voltage on the DC output. For every possible voltage on the scale of 0-125% rated voltage, the loaded IU table holds a current value, which can be any value between 0 and the rated current.